Herunterladen Inhalt Inhalt Diese Seite drucken

Troubleshooting Guide - MasterCool 55800 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 6
Replacing the Sensor
To replace the sensor, firmly grasp the flexible probe near the end with one hand and use the
other hand to unscrew the nozzle portion from the threaded probe tip in a counter-clockwise di-
rection. Next, remove the metal washer, the rubber washer and the sensor in that order. Observe
the orientation of the key on the sensor being removed. It would be advisable to replace the
filter inside the nozzle at the same time. To remove the microfiltration membrane from inside the
nozzle, the assistance of a long thin object such as an o-ring pick or equivalent will be required.
When installing a new sensor, orient the tab key on the sensor with the keys slot on the probe,
making certain that the 3 pins of the sensor are inserted into the appropriate pins inside the
probe end. Verify that the sensor has been properly inserted into the probe. Next install the
rubber washer as shown in Figure 2, followed by the metal washer. Insert new filter and screw
on the nozzle while firmly holding the probe end with the other hand. DO NOT USE ANY TOOLS in
tightening the tip assembly, firmly hand tighten only.
Leak Test Vial
A leak test vial (Figure 3) is provided with each leak detector for the convenience
of the user. To use the vial, unscrew the cap and place the sensor near the opening
of the vial as shown in Figure 3. Do not remove any material from the Test Vial. At
the conclusion of the vial test, replace the cap and tighten firmly.
QUICK START TIPS ON HOW TO FIND LEAKS
A sudden whipping action of the probe or blowing into the sensor may
cause the leak detector to false alarm. Electronic sensors are incorpo-
rated into the product to detect such activity and to reduce and minimize
such false alarm. In the event such interference is detected, the LCD will
display a flashing INT (interference) legend momentarily then normal
operation can resume after the READY legend appears.
1. When starting the search for leaks, without a general knowledge of the magnitude of the leak,
set the instrument sensitivity to LOW. The LOW sensitivity will enable the unit to locate medium
as well as large size leaks.
2. Slowly move the probe approximately 3/8 inch (9 mm) above the areas of suspected leaking.
Move the probe past the leak to allow the probe to clear if a leak is detected. Do not hold the
probe at the site of a leak. Return the probe to the same area where a leak was detected for
verification.
3. In the event no leaks were found with a LOW sensitivity setting, increase the sensitivity to
MEDIUM and repeat step (2) above.
4. For locating extremely small leaks 0.1 oz/yr (2.8 grams) or less, the HIGH sensitivity scale
should be used. Due to the extremely small leak size, the tip of the probe should be moved as
close to the surface as possible (without physically touching any object). Physical touching
of an object will be detected by the probe sensors resulting in a brief INT (interruption) in the
detector.
5. When probing for a leak in a contaminated environment, the sensor will detect the
contaminated area and will automatically adjust for the new conditions. The LCD will display
CA (contaminated atmosphere) on the screen, after going through the re-calibration process
which will take approximately one minute to complete. During the re-calibration process, all
vertical bars become lighted and sequence down until all bars are extinguished. At that point,
the READY light comes on and the leak detection process can continue.

TROUBLESHOOTING GUIDE

With the exception of batteries and the sensor, the internal parts of the 55800 are not field
serviceable. If you experience an issue with the product, refer to the troubleshooting table
below.
ISSUE
The unit will not power up
The unit powers up briefly then
shuts OFF
4
Probe
PROBABLE CAUSE
1) Batteries are not installed Install new batteries. Follow
2) Batteries are installed
improperly
Batteries are very weak
www.mastercool.com
-
Rubber
Metal
Key Sensor
Washer
Washer Filter
SOLUTION
instructions on page 3.
Check battery installation
polarity on page 3.
Install new batteries. Follow
instructions on page 3.
Cover
Screw for
removing
or securing
the battery
cover
+
Nozzle
Figure 1
Figure 2
Figure 3

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis