MAINTENANCE AND TROUBLESHOOTING
The machine does not need any particular maintenance apart from the periodic 
cleaning of the heating unit which is necessary to avoid the deposit of substances 
inside the seat of the biological indicators.  
For cleaning of the incubator is sufficient to refer to the most elementary rules of 
hygiene health provided in such circumstances. 
 First turn off the machine and pull the plug from the outlet 
 If the machine has just been used, waiting for cooling. 
 Use a sponge slightly moistened with soapy water and wipe dry with a dry and 
clean  
 Do  not  use  detergents  which  could  alter  in  irreversibile  way  the  appearance  of 
plastic parts of the machine. 
 Do  not  spray  directly  with  nebulizer  on  the  machine,  risking  damage  to  the 
electrical equipment inside.
 
DO NOT DO MAINTENANCE OR INSIDE REPAIR TO THE DEVICE. 
DISMANTLE THE DEVICE MEANS BREAKING IT.
 
 
Defect 
Pushing START button 
the incubator does not 
work 
The LED is turned on 
during the test 
 
After about 15 minutes 
of operating time the 
led switch on again 
 
 
 
Cause 
Intervention 
There is no power 
Control plug fit correctly in autoclave socket  
supply 
Resistance is  
Contact maker for technical assistance 
burned 
Push start button; led switch off,remove tubes 
(biological testing is to be considered unsuccessful) 
and leave it on for about 15 minutes.  
If led don't switch on automatically, means that the 
Electrical energy 
previous stopping was due to a temporary lack of 
interruption 
electrical energy.  
Otherwise if device switch on, means that the 
resistance is short‐circuit and security locks.  
Contact maker for technical assistance 
The heating unit 
Try to repeat the cycle. 
not arrived in 
In the case of persistent problem contact maker for 
temperature 
technical assistance 
Inside failure 
 
13
 