The Menu Map
PIPETTE
RPIPET
STEPPER
DILUTE
PROGRAM
OPTIONS
CALIBRATE
SERVICE
LANGUAGE
NAME
POWER
BUZZER
VERSION
BACKLIGHT
Pipette (Forward technique)
Choose the PIPETTE function as described above.
Select the pipetting volume simply by pressing the scroll key up or down. Accept the volume with
OK. Optionally press SPEED and the speed in starts to blink. Select the speed with scroll key and
accept with OK. Select the speed out with scroll key and accept with OK.
1. Dip the tip under the surface of the liquid in the reservoir and press the trigger. The liquid is
drawn in to the tip.
2. Wait until the liquid is not moving in the tip and withdraw the tip from the liquid touching it
against the edge of the reservoir to remove excess liquid.
3. To deliver the liquid press the trigger and hold it down. The blow-out is also included to
empty the tip.
4. Release the trigger to return to the ready position.
If necessary, change the tip and continue pipetting.
Rpipet (Reverse & Repetitive)
With RPIPET function both reverse and repetitive techniques are possible.
Choose the RPIPET function as described above.
Select the pipetting volume simply by pressing the scroll key up or down. Accept the volume with
OK. Press SPEED and the speed in starts to blink. Select the speed with scroll key and accept with
OK. Select the speed out with scroll key and accept with OK.
Reverse technique
The reverse technique is suitable for dispensing liquids that have a high viscosity or a tendency to
foam easily. The technique is also recommended for dispensing very small volumes.
Fill a clean reagent reservoir with the liquid to be dispensed.
1. Dip the tip under the surface of the liquid in the reservoir and press the trigger. This action
will fill the tip.
2. Withdraw the tip from the liquid touching it against the edge of the reservoir to remove
excess liquid.
3. Deliver the preset volume by shortly pressing the trigger. Some liquid will remain in the
tip and this should not be included in the delivery. A text BLOWOUT is displayed.
4. To empty the tip, press trigger again.