General information

Electrical safety

– The device must not be connected to the mains socket
if the mains plug or cables are damaged. The mains
power cable must be replaced with a special power ca-
ble available from the manufacturer or its customer
services.
Order number: see page 11.
– This unit is only to operated with an alternating voltage
of 230-240 V with 50 Hz.
– The device must only be connected to a socket with
correctly installed earthing contacts.
– The system should protected using a residual-current
circuit breaker (residual-current device) with a maxi-
mum rated residual current of 30 mA.
– Before connecting the device, please make sure that
the power supply corresponds to the device's power
supply data.
Notes on installing and operating the device
– The device is intended for installation in rooms or facil-
ities that are dry and not exposed to splashing or drip-
ping water. It must not be installed in rooms where the
maximum relative humidity is over 60%.
– Please contact customer service for advice if the de-
vice is intended for use at altitudes of 2000m above
sea level or above.
– The installation location must be easy to access for
maintenance work.
– The unit must be completely assembled and installed
so that no UV-C radiation can escape.
– Please adhere to the prescribed maintenance inter-
vals.
– This device may be used by children over the age of 8,
persons with reduced physical, sensory or mental ca-
pabilities and persons who have little knowledge or ex-
perience of the device; however, they must be super-
vised at all times. Do not allow children to play with the
device. Cleaning and user maintenance tasks must not
be carried out by children.
– When the installation conditions are no longer fulfilled
(e.g. due to changes to the operating environment), the
unit must not be operated any more.
– A socket with earthing contact must be provided by the
customer near to the unit.
– The installation location must be selected so that de-
germed air can equally flow through the room.
– Avoid positioning in the immediate vicinity of seating
areas or workplaces as well as beds (hospital wards)
due to the annoying flow of air.
– The circulation of air should not be interrupted by bar-
riers such as room partitions, large plants, etc.
6/28 - STERIBASE 300
– When installing on a floor with carpet, more dust parti-
cles will be extracted that overload the filter additional-
ly. We recommend the operation on a smooth floor or
when a smooth subframe is placed underneath (e.g a
panel).
– With longer periods of absence or during operating in-
terferences, pull out the mains plug to disconnect the
device from the power supply. Even when the mains
switch is located in the OFF position, the unit is not fully
separated from the mains supply. We recommend
turning the device off at the mains plug if the device is
not used for temporary periods (e.g. overnight).
– The following minimum clearances must be observed
for the selection of the installation location.
a
2 500 mm
b
500 mm
h = 1640 mm
1007835-01
h
a
b
b
230 - 240 V
40325 / 1
12/2014