Replacing the UV-C lamps

DANGER!
Risk of eye and skin damage due to UV-C ra-
diation!
Even when radiation periods are short, UV-C ra-
diation can severely damage the skin and eyes!
Direct irradiation of eyes and skin must be strict-
ly avoided.
– Disconnect the device from the power supply
before carrying out any maintenance work
and before opening the unit.
– Never operate UV-C lamps outside the device
or when it is open!
1
40328 / 1
6 x
40330 / 1
CAUTION!
Using unsuitable lamps can eliminate the
effectiveness of the unit and void the CE
conformity!
– Use genuine UV-C lamps only.
CAUTION!
Risk of damage to the device!
UV-C lamps are sensitive components and may
be easily damaged.
– Do not touch the lamps with bare hands..
Wear protective gloves.
– Take care when touching and installing the
lamps. Do not bend the lamps. Do not rest the
lamps on the pins.
2
1
1
2
Installation and operation
40331 / 2