Example (HF58B, HF58B-r, HF59B, HFW59B, HFE59B: Attenuator connected and
the "Adapter" switch set for "Attenuator -20dB):
Displayed value:
100 µW/m²
HFE35C, HFW35C:
The value from the display must be multiplied with a factor of 100 to get the real
field strength:
Example : Displayed value:
100 µW/m²

Technical data:

Transmission band:
Return loss:
Diagram:
0 dB
-5 dB
-10 dB
-15 dB
-20 dB
-25 dB
-30 dB
-35 dB
-40 dB
-45 dB
-50 dB
0 GHz 1 GHz 2 GHz 3 GHz 4 GHz 5 GHz 6 GHz 7 GHz 8 GHz 9 GHz 10 GHz
4
© by GIGAHERTZ SOLUTIONS
R
100 µW/m² x 100 =10 000 µW/m² = 10,0 mW/m²
0 Hz (DC) < 8 Ohm
10 MHz - 10000 MHz = 20 dB ± 0,4 dB
10 MHz - 6000 MHz > 20 dB
6000 MHz - 10000 MHz > 8 dB
Transmission
True (precise) value:
100 µW/m²
True (precise) value:
Return loss
®
GmbH, D-90579 Langenzenn, www.gigahertz-solutions.de