CAUTION:Risk of injury from tripping or caused by the device
falling
A compressor placed in a poor position or for which the cables are
routed untidily poses a risk of injury.
Route the power cord in such a way that it does not present a trip-
ping hazard and people cannot become entangled in it.
Make sure that children are not able to pull the compressor down
by the power cord or connection tubing.
CAUTION:Quality of treatment impaired by electromagnetic
interference
Electrical devices can cause electromagnetic interference. Interfer-
ence can impair the function of the devices and thus also the effec-
tiveness of the treatment.
Do not place the INQUA device immediately beside or on top of
other devices.
Keep at a minimum distance of 30 cm from portable wireless
communication devices (including accessories therefore, such as
antenna cables or external antennas).
NOTICE:Risk of damage to the compressor due to overheating
and dust
The compressor is fitted with a fan to ensure that the motor is cooled
constantly. If the fan is not able to cool the motor sufficiently, the
motor may overheat and be damaged.
Only operate the compressor in locations with low dust contami-
nation and an adequate air supply. The ventilation slits of the
compressor must always remain unobstructed.
Never cover the compressor while it is in use.
Do not run the compressor on carpets or in very dusty environ-
ments, e.g. under the bed, in workshops etc.
Observe the ambient conditions specified in
section "Environmental conditions for operation", page 33.
The compressor is only intended for indoor use, it is not intended for
outdoor use.