Herunterladen Inhalt Inhalt Diese Seite drucken

Seca 984 Bedienungsanleitung Seite 27

Vorschau ausblenden Andere Handbücher für 984:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
– You can test the limit value function by
lifting the bed so that the measured val-
ue fails to reach the set threshold. The
alarm sounds. If you release the bed
again, the alarm goes silent.
– In order to switch off the limit value
function, press the FUNC key again.
The arrow indicator for limit value func-
tion goes out.
Important note:
if the scale has been switched off, the limit value function must be reactivated as
described above.
Weighing with pre-Tare
If the pre-Tare function is activated, a stored weight value is deducted from the currently
measured weight. This function is useful, for example, if the weight of a bed is known and
is not to be displayed. As a result, the patient can remain in bed throughout the entire
weighing process.
Tip:
Make a note of their tare weight on the beds used for weighing - on an adhesive label, for
example.
– Prepare for measurement as described
in the Section entitled "Preparation for
measurement" on page 20, but do not
yet push the measuring feet under the
bed castors.
– Press the START key with no load on
the scale.
,
se(a
8. 8 . 8 . 8 . 8 .
secutively in the display. The scale is
then automatically set to zero and ready
for operation.
– Press the FUNC key.The last function
used appears in the display.
– Use the arrow keys to select the pre-
Tare function, FU 3.
Activate the function by pressing the
FUNC key again.
In the display you will see the tare value
from the memory.
– You can use the arrow keys to adjust
the tare value in increments according
to the graduations displayed.
Model 984
and
appear con-
0. 0
27
GB

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis