Herunterladen Inhalt Inhalt Diese Seite drucken

Hameg Instruments LCR-Bridge HM8118 Handbuch Seite 33

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

200 kHz LCR-Bridge HM8118
All data valid at 23 °C after 30 minutes warm-up.
Conditions
Test signal voltage:
1 V
Open and short corrections performed
Measurement time:
SLOW
Display
Measurement modes:
Auto, L-Q, L-R, C-D, C-R, R-Q, Z-Θ, Y-Θ,
R-X, G-B, N-Θ, M
Equivalent circuits:
Auto, Series or Parallel
Parameters displayed:
Value, Deviation or % Deviation
Averaging:
2...99 measurements
Accuracy
Primary Parameters:
Basic accuracy
(Test voltage: 1.0 V,
measurement SLOW/MEDIUM,
autoranging mode, constant voltage OFF,
bias off). For FAST mode double the basic
accuracy values
Impedance: 100 MΩ
0.2% + |Z|/1.5GΩ
4 MΩ
1 MΩ
0.05% +
|Z|/2GΩ
25 kΩ
100 Ω
0.1% + 1mΩ/|Z|
2.5 Ω
0.3% + 1mΩ/|Z|
0.01mΩ
20 Hz
Secondary Parameters:
Basic accuracy D, Q:
±0.0001 @ f = 1 kHz
Phase angle:
±0.005° @ f = 1 kHz
Ranges
|Z|, R, X:
0.01 mΩ...100 MΩ
|Y|, G, B:
10 nS...1000 S
C:
0.01 pF...100 mF
L:
10 nH...100 kH
D:
0.0001...9.9999
Q:
0.1...9999.9
-180...+180°
Θ:
∆:
-999.99...999.99 %
M:
1 µH...100 H
N:
0.95...500
Measurement conditions and functions
Test frequency:
20 Hz...200 kHz (69 steps)
Frequency accuracy:
±100 ppm
AC test signal level:
50 mV
Resolution:
10 mV
Drive level accuracy:
±(5 % + 5 mV)
Internal Bias Voltage:
0...+5.00 V
Resolution:
10 mV
External Bias Voltage:
0...+40 V
Internal Bias Current:
0...+200 mA
Resolution:
1 mA
Range Selection:
Auto and Hold
Trigger:
Continuous, manual or external via
interface, Binning Interface or Trigger
Input
Trigger delay time:
0...999 ms in 1 ms steps
HM8118E/250112 · C&E · Subject to change without notice · © HAMEG Instruments GmbH
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · info@hameg.com
0.5% +
|Z|/100MΩ
0.1% +
|Z|/1,5GΩ
0.2% +
|Z|/100MΩ
0.5% +
5mΩ/|Z|
+
0.2% +
|Z|/10MΩ
2mΩ/|Z|
0.5% +
2mΩ/|Z|
1 kHz
10 kHz
100 kHz
...1.5 V
rms
rms
rms
dc
Specifications
(fused 0.5 A)
dc
Measurement time (f ≥1 kHz)
FAST
70 ms
MEDIUM
125 ms
SLOW
0.7 s
Other Instrument Functions
Test signal level monitor:
Voltage, current
Error Correction:
Open, Short, Load
Save/Recall:
9 instrument settings
Front-end Protection:
V
max
(1 Joule of stored energy)
Low Potential and
Low Current Guarding:
Ground, Driven Guard or Auto (fused)
Constant Voltage Mode (25 Ω source)
Temperature effects:
R, L or C:
±5 ppm/°C
Interface:
USB/RS-232 (HO820),
IEEE-488 (GPIB) (optional)
Safety Class:
Safety Class I (EN61010-1)
Power supply:
110...230 V ±10 %, 50...60 Hz, CAT II
Power consumption:
approx. 20 W
Operating temperature:
+5...+40 °C
Storage temperature:
-20...+70 °C
Rel. humidity:
5...80 % (non condensing)
Dimensions (W x H x D):
285 x 75 x 365 mm
Weight:
approx. 4 kg
Accessories supplied: Line cord, Operating manual, HZ184 4 Terminal Kelvin
Test Cable and HZ188 4 Terminal SMD Component Test Fixture, CD
Recommended accessories:
HO118
Binning Interface
HO880
IEEE-488 (GPIB) Interface (galvanically isolated)
HZ13
Interface cable (USB) 1.8 m
HZ14
Interface cable (serial) 1:1
HZ33
Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34
Test cable 50 Ω, BNC/BNC, 1 m
HZ42
19" Rackmount kit 2RU
HZ72
GPIB-Cable 2 m
HZ181
4 Terminal Test Fixture including Shorting Plate
HZ186
4 Terminal Transformer Test Cable
®
· DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
S p e c i f i c a t i o n s
2 /C @ V
<200 V, C in Farads
< ZX
max
Subject to change without notice
33

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis