Aqualytic SD 320 Con Bedienungsanleitung Seite 43

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Menu
Parameter
rEAd Logg: Review stored measurements,
see Chapter 8.1! Manual recording ("Func-Stor")
Set Configuration: General settings
SD 320_1 03/2013
Values
Meaning
Input: Selection of the measurement
Cond
Conductivity
rES:
Specific resistance
SAL
Salt content/salinity
TDS
Total Dissolved Solids
TDS measurement: Conversion factor
(only with Inp = TDS)
0.40 - 1.00
Conversion factor for TDS measurement
Cell Range: Adjustment of the cell constant: Cell
constant range
0.01
Purest water, electrodes with K ~ 0.01
0.1
Purest water, electrodes with K ~ 0.1
1
Standard electrode in the standard delivery
contents, e.g.
SET1 K = 0.55
SET2 K = 0.40
10
Electrode with K=10
Cell Factor: Adjustment of the cell constant:
Multiplication factor
0.3800 -
Multiplication factor of the cell constant:
1.5000
Cell constant CELL = CELL Range
Factor
t-Input: Selection of the temperature input
NTC
NTC 10k sensor (LC 12)
Pt
Pt1000 sensor (LC 16)
Range: Selection of the display range (conductivity,
spec. resistance or TDS)
Auto
Automatic range selection
e.g.
Example for CELL rAng 1 and InP Cond:
0.0 ... 500.0
for others, see chap. 6.3 - 6.5
µS/cm
...
0 - 1000 mS/cm
Example for CELL rAng 1 and InP Cond:
for others, see chap. 6.3 - 6.5
**
.
CELL
.
43