Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Electrodes
number of electrodes
electrode material
Shaft material
Dimensions
shaft diameter
shaft length
diameter of connection head
total length of the sensor
cable length
Immersion depth
Pressure resistance
Connections
cable, plug
measuring cell
Cell constant
Measuring range
Application range with LF 330/340
Temperature measurement
thermistor type
thermistor material
temperature range
response time of thermistor
sensor accuracy
Storage medium
66

Technical data

4
graphite
Epoxy, black
15.3 mm
120
mm
21.7 mm
162.5 mm
1.5 m (cable is exchangeable by
WTW-Service)
min. 36 mm
max. total length incl. cable ( ϑ < 80°C)
max. 120 mm ( ϑ < 100°C)
2 bar
protection system IP67, 8-pin plug
protection system IP68 (20 m)
± 1.5 %
-1
0.475 cm
1 µS/cm ... 2 S/cm
1 µS/cm ... 500 mS/cm
integrated NTC (30 kOhm/ 25°C)
graphite
-5 .. +100°C
t
< 20 sec.
99
± 0,2 K
air

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Tetracon 325Lf 330/set

Inhaltsverzeichnis