Conditions for use and storage
User weight:
Minimum: 15 kg, maximum: 250 kg
Temperature:
Minimum: -20 °C, maximum: 55 °C
Use at high altitude
Notice! Vicair B.V. recommends not to use
the cushion above 2400 m (8000 ft) above
sea level or below 50 m (164 ft) below sea
level. Because each SmartCell is filled with
air, the cushion may behave differently when
used at high altitude or in an airplane. Under
such circumstances, the cushion's firmness
can be adjusted by removing a few cells from
each compartment.
Air humidity:
Minimum: 0%, maximum 100%
Daily fluff-up
Vicair B.V. recommends to fluff-up your Vicair
cushion before use on a daily basis.
Use of Vicair seat cushions in motor vehicles
Warning! No testing data is currently available
about the safe use of Vicair wheelchair cushions
in motor vehicles. Therefore we do not
encourage using the wheelchair cushion in
motor vehicles. As an alternative for your Vicair
wheelchair cushion, the Vicair AllRounder
(which can be fastened to the user's body) can
be used if you require additional protection
during transportation.
Re-use - Transferring the product to another
user
The product is suitable for repeated use.
The number of times it can be used depends on
how often and in which way the product is used.
The product has to be thoroughly cleaned and
disinfected, before another person is allowed to
use the it. See chapter 'Maintenance and
cleaning' for instructions.
Notice! Make sure the product is correctly
adjusted to the other person.
6
Replacing the cushion cover
Under normal use, Vicair B.V. recommends to
replace the cushion cover every year.
Replacement covers can be purchased
separately.
Disposal
The disposal and recycling of used devices and
packaging must comply with the applicable legal
regulation. Ensure the product is cleaned prior
to disposal to avoid any risk of contamination.

MAINTENANCE AND CLEANING

Warning! Every three months, check the product
for wear and deflated cells. In case of serious
damage, or a large number of deflated cells, do
not continue to use the product. Consult your
dealer.
Caution! Do not steam clean or autoclave the
product. It cannot withstand excessive heat or
pressure. When the given cleaning instructions
are not followed, it is very likely that the product
will age more rapid.
Caution! Do not use aggressive or abrasive
cleaners, or sharp objects to clean the cushion.
Cleaning the Vicair Adjuster O2
Cleaning the Vicair Adjuster O2 cushion
cover
The cushion cover can be machine washed.
Follow the washing instructions on the label
attached to the inside of the cover.
Cleaning the Vicair Adjuster O2 cushion
Your Vicair O2 cushion can be machine
washed at 60°C using a gentle cycle and mild
detergent. You do not need to remove the
SmartCells before putting the cushion into
the washing machine. Do not use any
E N