Tabelle 16 Busparameter des Umformerblocks (Fortsetzung)
Parameter
Temperature
SECONDARY_VALUE_
UNIT_2
TEMP_SENSOR_COMP
TEMP_SENSOR_MAN_
VALUE
TEMP_SENSOR_CALIB
TEMP_SENSOR_CALIB_
MAN_VALUE
TEMP_SENSOR_TYPE
Description
Degree C or degree F. Changes
the unit of temperature being
displayed and transmitted.
Indicates manual or automatic
temperature compensation.
Turns automatic pH sensor
temp. compensation on and off.
Temperature value used in
manual temp. compensation
mode.
The constant temp. value used
to calculate pH in the manual
temp. compensation mode
Indicates manual or automatic
mode of temperature
measurement for calibration.
Temperature value used in
manual temp. compensation for
calibration.
Type of temperature sensor.
The value entered must
correspond to the temp.
element in the pH sensor being
used.
Data
R/W Bytes
Range
type
1001 = °C
R/W 2
uns16
1002 = °F
0 = Automatic
R/W 1
uns8
1 = Manual
R/W 4
float
-20–200 °C
0 = Automatic
R/W 1
uns8
1 = Manual
R/W 4
float
-20 – 200 °C
128 = Pt100
200 = Pt1000
R/W 2
uns16
1000 = NTC300
1001 = NTC8.55
1
2