GUIDANCE AND MANUFACTURER'S DECLARATION-ELECTROMAGNETIC IMMUNITY
Icare TA01i is intended for use in the electromagnetic enviroment specified below. The customers
or users of Icare TA01i should assurre that it is used in such enviroment
Immunity test
Radiated RF IEC
61000-4-3
Conducted RF
IEC 61000-4-6
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the Icare TA01i is used exceeds the applicable
RF compliance level above, the Icare TA01i should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
or relocating the Icare TA01i.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
IEC 60601
Compliance
Test level
level
3 V/m
3V/m
80MHz to 2.5 GHz
3Vrms
NOT
150 kHz to 80 MHz
APPLICABLE
Electromagnetic environment-
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of
the IcareTA01i, including cables,
than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation
distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2 5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey should
be less than the compliance level in
each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
English
9