Herunterladen Inhalt Inhalt Diese Seite drucken

Guidance And Manufacturer's Declaration-Electromagnetic Emissions; Guidance And Manufacturer's Declaration-Electromagnetic Immunity - Icare ta01i Gebrauchsanweisung

Tonometer
Vorschau ausblenden Andere Handbücher für ta01i:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 16
Icare tonometer TA01i manual

GUIDANCE AND MANUFACTURER'S DECLARATION-ELECTROMAGNETIC EMISSIONS

TA01i is intended for use in the electromagnetic environment specified below. The user of the
TA01i should assure that it is used in such an environment.
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
flickering emissions
IEC 61000-3-3

GUIDANCE AND MANUFACTURER'S DECLARATION-ELECTROMAGNETIC IMMUNITY

Icare TA01i is intended for use in the electromagnetic enviroment specified below. The customers
or users of Icare TA01i should assurre that it is used in such enviroment
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
Transients/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruption and
voltage variations on
power supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
8
Group 1
Class B
NOT APPLICABLE
NOT APPLICABLE
IEC 60601
Compliance
Test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
±2 kV for power
NOT
supply lines
APPLICABLE
±1 kV for input/
output lines
±1 kV for line(s) to
NOT
line(s)
APPLICABLE
±2 kV for line(s) to
earth
<5 % UT
NOT
(>95 % dip in UT)
APPLICABLE
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
3 A/m
Icare TA01i is battery operated and use RF
energy only for its internal function. Therefore,
its RF emissions are low and are not likely to
cause any interference in nearby equipment.
Icare TA01i is suitable for use in all
establishments, including domestic
establishments and those directly connected to
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic environment-
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis