rated voltage and current of the ZA-TEC ZA 6-2.
-
Check the insulation of the safety measuring leads. If the insulation is dam-
aged in any way, do not use the leads.
-
Check the continuity of the safety measuring leads. If the conductor in the
safety measuring lead is interrupted, do not use the leads.
-
Before selecting another function with the rotating switch  , always discon-
nect the safety measuring leads from the measuring point.
-
Sources of strong current in the vicinity of the ZA-TEC ZA 6-2 may cause
unstable or incorrect readings.
8.2 Voltage and current measurement
Always observe the maximum voltage to earth potential!
Electrical hazard!
The maximum voltage which may be applied to the sockets
-
COM socket 
-
socket for V, Ω,
-
socket for 10 A range L
of the ZA-TEC ZA 6-2 to earth is 1000 V CAT III.
8.2.1
Voltage measurement
-
With the rotating switch  , select the desired function ( ,
ZA-TEC ZA 6-2.
-
Connect the black safety measuring lead to the COM socket  on the
ZA-TEC ZA 6-2.
-
Connect the red safety measuring lead to the socket  on the
ZA-TEC ZA 6-2.
-
Connect the safety measuring leads to the measuring points. Read the meas-
urement value displayed in the digital display  of the ZA-TEC ZA 6-2.
See fig. 2:
See fig. 3:
8.2.2

Current measurement

-
With the rotating switch  , select the desired range and function (A AC/DC
or µA AC/DC) on the ZA-TEC ZA 6-2.
-
Connect the black safety measuring lead to the COM socket  on the
ZA-TEC ZA 6-2.
-
Connect the red safety measuring lead to the jack for the A range L (up to
10 A AC/DC) of the ZA-TEC ZA 6-2.
-
Connect the safety measuring leads to the measuring points. Read the
measurement value displayed in the digital display  of the ZA-TEC ZA 6-2.
See fig. 4:
See fig. 5:
8.3 Resistance measurement
-
With the rotating switch  , select the desired function (Ω,
ZA-TEC ZA 6-2.
-
Connect the black safety measuring lead to the COM socket  on the
ZA-TEC ZA 6-2.
-
Connect the red safety measuring lead to the socket  on the
ZA-TEC ZA 6-2.
-
Connect the safety measuring leads to the measuring points. Read the meas-
urement value displayed in the digital display  of the ZA-TEC ZA 6-2.
See fig. 6:

8.4 Diode testing

-
With the rotating switch  , select the desired function (
ZA-TEC ZA 6-2.
-
Using the blue key  on the ZA-TEC ZA 6-2, switch to diode testing (
-
Connect the black safety measuring lead to the COM socket  on the
ZA-TEC ZA 6-2.
-
Connect the red safety measuring lead to the socket  on the
ZA-TEC ZA 6-2.
-
Contact the diode connections with the safety measuring leads and read the
measurement value displayed in the digital display  of the ZA-TEC ZA 6-2.
-
For a normal silicone diode located in flow direction, the flow voltage
between 0.4 V and 0.8 V is displayed. If "000" appears in the display, there
may be a short circuit in the diode.
-
If no forward voltage is detected, first check the polarity of the diode. If still
no forward voltage is displayed, the forward voltage of the diode is beyond
the measuring limits.
See fig. 7:
11/ 2016
, Hz 
DC-voltage measurement
AC-voltage measurement
DC-current measurement
AC-current measurement
Resistance measurement
Diode testing
ZA-TEC ZA 6-2
,
) on the
) on the
,
) on the
).
24