6. Packaging / 7. Starting, operation
6. Packaging
Has everything been supplied?
Check the scope of supply:
Completely assembled pressure switches; with flush version PS-20 including pre-assembled
■
sealings and protection cap.
Inspect the pressure switch for possible damage during transportation. Should there be any
■
obvious damage, inform the transport company and WIKA without delay.
Keep the packaging, as it offers optimal protection during transportation (e.g. changing installation
■
location, shipment for repair).
Ensure that the pressure connection thread and the connection contacts will not be damaged.
■
In order to protect the diaphragm, the pressure connection of the instrument PS-21 is provided with a special
protection cap.
Remove this protection cap only just before installing the pressure switch in order to prevent any
■
damage to the diaphragm or the thread.
Keep the protection cap of the pressure connection thread and the diaphragm for later storage or
■
transport.
Mount the protection cap when removing and transporting the instrument.
■
7. Starting, operation
Required tools: wrench (flats 27 or 41), screw driver
Diaphragm test for your safety
It is necessary that before starting the pressure transmitter you test the diaphragm visually, as this is a safety-
relevant component.
!
Pay attention to any liquid leaking out, for this points to a diaphragm damage.
■
Check the diaphragm visually for any damage (PS-21).
■
Use the pressure switch only if the diaphragm is undamaged.
■
Warning
Use the pressure switch only if it is in a faultless condition as far as the safety-relevant
■
features are concerned.
6
WIKA Operating instructions / Betriebsanleitung / Mode d'emploi
GB
PS-20, PS-21