390
3.7
Adjusting the balance using its internal reference weight
1.
Ensure the balance has reached its operating temperature, is levelled and set to the weighing
mode (turn the balance off and on by pressing [ON/OFF] twice).
2.
Wipe left > (Settings) > Adjustment > Adjustment mode > Internal or Automatic >
{Home}.
3.
Wipe left > (Adjustment) > {Start}. The balance adjusts itself using its internal reference
weight displaying "Adjustment running...". As soon as it is finished, it will display "Adjustment
done.".
3.8
Adjusting the draft shield
1.
Ensure the balance is levelled and set to the service mode according to chapter 3.3.
2.
Wipe left > (MSP) > {Adjust draft shield}. The balance adjusts the draft shield
automatically displaying "Draft shield adjustment running ...". When this is done it displays the
info window "Adjustment done".
3.9
Adjusting the bubble level
1.
Ensure the balance is set to the service mode according to chapter 3.3.
2.
Wipe left > (Levelling) > {StartAdj}.
3.
Grasp the scale from below, lift it carefully and turn it so that the air bubble goes completely
around once along the edge in a circle.
Precisa Gravimetrics AG
Adjustments
70