Herunterladen Inhalt Inhalt Diese Seite drucken

Accessories Supplied; Accessorie Hz560 - Hameg Instruments Hm5012-2 Handbuch

Spektrumanalysatoren
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Specifications
5012-
/ 5014-
2
2
Frequency
Frequency Range:
Stability:
Ageing:
Display Resolution (Readout):
Center Frequency Range:
Frequency Generation:
Span:
Zero span and 1MHz to 1000MHz in steps of 1-2-5
Marker:
Resolution (Frequency)
Resolution (Amplitude)
Resolution Bandwidth, RBW (6dB):
Video Bandwidth, VBW:
Sweep time (autom. Selection):
Amplitude (Marker related)
Measurement Range:
Scaling:
Display Range:
Frequency response (@10dB Attn., Zero Span, RBW 1MHz,
-20dBm Signal):
Display CRT, Units:
Input Attenuator Range:
Attenuator Accuracy rel. to 10dB:
Maximum Safe Input Level (continuous):
Attenuator setting 40dB:
Attenuator setting 0dB:
DC max:
Reference Level Variation:
Reference Level Accuracy (rel. to 500MHz, 10dB Attn., Zero Span,
RBW 1MHz):
Min. Average Noise Level:
Intermodulation (3rd Order):
(2 Signals, –27 dBm each, Frequency distance>3MHz)
Harmonic Distortion
(2nd at input level -27dBm, ATTN 0dB, Frequency distance>3MHz)
Resolution Bandwidth Amplitude Error (rel. to RBW 1MHz,
Zero Span):
Digitising Error:
±1 Digit (0.4dB) @ 10dB/div. (Average mode)

Accessories supplied:

Optional Accessories: Opto-Interface HZ70, Sniffer Probes HZ530,
Transient Limiter HZ560, Line Impedance Stabilization Network HM6050-2
Accessory
HZ560 Transient-Limiter
The transient limiter HZ560
HZ560
HZ560 protects
HZ560
HZ560
a spectrum analyzer input from damage
caused by high level transients from line
impedance stabilization networks during
EMC testing for conducted emissions.
The transient limiter can withstand inputs
as high as 2W of average power. The
built in high pass filter helps reduce 50/
60Hz line feed through. The use of this
limiter is strongly recommended when
measuring with a Line Impedance Stabi-
lization Network.
28
(Reference Temperature 23°C ±2°C)
0.15 MHz to 1050MHz
±5ppm
±1ppm/year
1kHz (6½ -Digit)
0 MHz - 1050 MHz
TCXO with DDS
1KHz, 6½ -Digit
0.4dB, 3½ -Digit
9kHz, 120kHz and 1MHz
40ms, 320ms, 1s*
–100dBm to +10dBm
10dB/div., 5dB/div.
80dB (@ 10dB/div.)
40dB (@ 5dB/div.)
8 x10, logarithmic, dBm
0 - 40dB, 10dB steps
+20dBm (0,1W)
+10dBm
-99.6dBm to+10dBm
–100dBm (9kHz RBW)
better than 75dBc
better than 75dBc
Instruction Manual, Power Cord, PC-Software on CDR
Frequency Range: 150kHz to 30MHz
Insertion loss:
Max. Input level:
Continuous 2W average (+33dBm)
Inputs / Outputs
Signal Input:
Input Impedance:
Tracking Generator Out (HM5014)
Output Impedance:
Test Signal Output
Output Impedance, Frequency, Level:
Probe Power:
VSWR:
Audio Output (Phone):
4kHz
Functions
Num. Keypad:
Rotary Dial:
Max. Hold Detection:
Quasi-Peak Detection:
Average:
±3dB
SAVE/RECALL
AM-Demodulator
LOCAL:
±1dB
Tracking Generator (HM5014-
Output Frequency Range:
±25 V
Output Level:
Frequency Response:
Spurious Outputs:
±1dB
General Data
Temperature Range:
Power Requirements:
Power Consumption:
Protective System:
±1dB
Dimensions:
Cabinet, Weight:
10dB (+1.5 / –0.5dB)
>90 dB below 1kHz
>50 dB over 100MHz
DC-voltage ±50Vdc
N-Connector
N-Connector
50Ω, 48MHz,
-30dBm ±2dB
6V (Near field probes)
ATTN ≥10dB
3.5 mm
Center Frequency,
Reference and Tracking Generator Level
Center Frequency, Reference and
Tracking Generator Level, Marker
Peak Detection
with software AS100E
Mean value measurement
10 set-up settings
RS-232 control disabled
only)
2
150kHz to 1050MHz
–50dBm to +1dBm
better than 20dBc
Operating: 10°C-40°C
100-240V~ ±10%, 50-60Hz
approx. 43W
Safety Class I (IEC 1010-1/EN61010-1)
W 285, H 125, D 380mm
Lockable tilt handle, approx.6kg
VSWR:
1.5:1 or better
Connectors:
BNC (input and output)
Dimensions:
Subject to change without notice
Subject to change without notice
50Ω
50Ω
BNC (F)
typ. 1.5 : 1
jack
Ear Phones
±2dB
W 32mm
H 67mm
D 32mm

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Hm5014-2

Inhaltsverzeichnis