2.1

Where to install

The transmitter can be installed above or below the pressure tapping point.
When measuring gases, we recommend that the transmitter be installed above
the pressure tapping point and that the pressure pipe be laid so it runs down to
the pressure tap. This will permit any condensation in the pipe to drain off so that it
cannot affect the measurement.
When measuring liquids, the transmitter should be installed below the pressure
tapping point and the pipe should be laid so it rises constantly up to the pressure
tap, thus enabling any gas trapped in the pipe to escape.
The point of installation should be easily accessible, preferably close to the
measuring point and should be free from vibration. The permitted ambient
temperature limits must not be exceeded. The transmitter must be protected
against direct heat sources.
Before installing the transmitter, compare the process data against the data on the
rating plate.
Keep the housing closed during the installation process.