Herunterladen Diese Seite drucken

Knick SE 630 Memosens Betriebsanleitung Seite 10

2-elektroden-leitfähigkeitssensor
Vorschau ausblenden Andere Handbücher für SE 630 Memosens:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1. Applications
The SE 630 Memosens® conductivity sensor is a 2-electrode sensor for measur-
ing low to medium conductivity values over a large pressure and temperature
range. The sensor consists of corrosion-resistant materials (graphite electrodes
and PES body). A temperature detector is integrated in the sensor.
Preferred fields of application are water, moderately polluted wastewaters,
process solutions with medium conductivities; also corrosive media.
2. Dimension Drawing
79
16
Polyethersulfone
3. Calibration / Adjustment
Remove sensor, clean, rinse, and dry it, then immerse it in a calibration solution.
Suitable calibration solution: ZU 0348 (12.88 µS/cm), ZU 0349 (1413 µS/cm).
Perform a product calibration if the sensor can or shall not be removed.
Observe user manual of meter. The cell constant of the sensor can be certified
by the manufacturer (ZU 0320). For that purpose, you must send in the sensor.
4. Application in Hazardous Locations
Serie
Type (1)
Explosion-
protected
SE6
30
(1) not Ex-relevant
II 1G Ex ia IIC T3/T4/T6
BVS 10 ATEX E 041 X
10
186
18
G 1;
Memosens
X
-
MS
W 36
Ambient temperature
Ta with temperature class
-20 °C ≤ Ta ≤ 135°C (T3)
-20 °C ≤ Ta ≤ 115°C (T4)
-20 °C ≤ Ta ≤ 65°C (T6)
0044
Memosens®
cap
SE 630-MS gray,
SE 630X-MS orange-red
Category
II 1 G

Werbung

loading

Diese Anleitung auch für:

Se630x-ms