Applications
Age: Micro: 2-3 yrs., Small: 3-5 yrs., Medium: 5-9yrs., Large: 9-young adult yrs.
Limb Description: All Models: Trans-radial & mid-length - long Trans-humeral
Preparation
Consider the following factors as you determine the most appropriate technology and
prosthesis design for your client:
1.
Does the user or caregiver understand the function and safety features of the
device?
2.
What other terminal devices will be used with this prosthesis?
3.
In what types of environments will the device(s) be used?
4.
In what positions will the person use the device?
5.
How will the user switch terminal devices? Will quick-disconnect be helpful?
6.
For what other activities might this device be helpful?
7.
Will the person benefit from additional training in the use of the device?
Installation
LITE TOUCH BIOMECHANICAL HANDS may be installed in any Fillauer TRS Omega
Wrist, or Fillauer LLC wrist units with a ½-20 thread. NOTE: Use with Economy Friction
Wrists is not recommended. Follow the instructions provided with the wrist unit for
best results.
Cabling
The pivoting "thumb" has been designed with a receiver for a 3/16" terminal ball. A
Binary Swivel, or triple swivel with a 3/16" terminal ball should be selected with a cable
connection that matches the cable used in the prosthesis. Cable routing should assure
a direct line of pull that minimizes bends in the cable which could result in excess cable
friction or failure .