Performing a Glucose Control Solution Test
Note:
- Gluceo ne® glucose control solution is sold separately.
- It is recommended that glucose control solution be stored at room temperature
20-25°C (68-77
) before testing.
For example:
If you use a normal glucose control solution, you must use the normal range printed
on the side of the blood glucose test strip vial.
Results that fall outside the expected range may indicate:
- Test not performed at room temperature (20-25°C).
- An error in the way you are doing the test.
- Expired or contaminated glucose control solution.
- A problem with the blood glucose test meter.
- Glucose control solution that is too warm or cold.
- Failure to discard the rst drop of glucose control solution and to wipe the bottle
tip clean.
- Failure to shake the glucose control solution adequately.
CAUTION
• If your glucose control solution falls out of range, do not perform a glucose test, please contact
your local representative.
• The Glucose Control Solution range is not the recommended range for your blood glucose level.
62
Kontroll-Lösung / Solution de contrôle /
Control Solution / Soluzione di controllo
(Normal / Normale)
Vor Gebrauch gut schütteln / Shake well before use
Bien agiter avant l'emploi / Agitare bene prima dell'uso
0483
4212003
METRADO GmbH | Auf Pfuhlst 1 | D-66589 Merchweiler | Tel. +49 6825 95 22 78 -0
Kontroll-Lösung / Solution de contrôle /
Control Solution / Soluzione di controllo
(Hoch / High / Élevée / Alto)
Vor Gebrauch gut schütteln / Shake well before use
Bien agiter avant l'emploi / Agitare bene prima dell'uso
0483
4212003
METRADO GmbH | Auf Pfuhlst 1 | D-66589 Merchweiler | Tel. +49 6825 95 22 78 -0