13. Heating and cooling times
13.7 CTB9100-225 micro calibration bath
Measuring conditions
All times refer to a Pt100 reference sensor with Ø 6 mm
The reference sensor is positioned centrally in the tank,
5 mm over the strainer insert
Distilled water
Heating up:
25 °C to 40 °C
40 °C to 50 °C
50 °C to 75 °C
75 °C to 90 °C
Silicone oil 5 CS
Heating up:
25 °C to 40 °C
40 °C to 50 °C
50 °C to 75 °C
75 °C to 100 °C
100 °C to 120 °C
Silicone oil 10 CS
Heating up:
25 °C to 40 °C
40 °C to 50 °C
50 °C to 75 °C
75 °C to 100 °C
100 °C to 125 °C
125 °C to 150 °C
150 °C to 165 °C
WIKA operating instructions, models CTD9100, CTB9100, CTM9100
Time
0:55 min
0:37 min
1:27 min
1:30 min
Time
0:51 min
0:16 min
0:54 min
1:13 min
1:35 min
Time
0:52 min
0:22 min
0:52 min
0:53 min
0:59 min
1:12 min
1:03 min
All times are transition times and do not account for the
required settling time
The measurements were carried out at a room tempera-
ture of approx. 23 °C, with the cover off the bath
Cooling:
90 °C to 75 °C
75 °C to 50 °C
50 °C to 40 °C
40 °C to 25 °C
Cooling:
120 °C to 100 °C
100 °C to 75 °C
75 °C to 50 °C
50 °C to 40 °C
40 °C to 25 °C
Cooling:
165 °C to 150 °C
150 °C to 125 °C
125 °C to 100 °C
100 °C to 75 °C
75 °C to 50 °C
50 °C to 40 °C
40 °C to 25 °C
Time
5:53 min
15:17 min
10:50 min
45:26 min
Time
3:27 min
5:55 min
10:00 min
7:02 min
34:28 min
Time
1:40 min
3:17 min
4:14 min
5:59 min
9:59 min
7:00 min
31:40 min
43