7. Operating elements of the calibrator/micro calibration bath
7.4.6 Setting and storing fixed set temperatures
In order to store the set temperature in the calibrator or micro
calibration bath, the appropriate set-point memory must be
opened.
GB
In calibration mode, press button P for approx. 5 secs., this
opens the main menu.
The upper PV display will show OPEr.
In the lower SV display, the LED SET will blink.
Operator menu OPEr
If button P is pressed again, this opens the groups level.
The upper PV display will show OPEr.
In the lower SV display, 'SP is shown and additionally LED
SET will blink.
'SP Group
If button P is pressed again, this opens the parameter level.
The upper PV display will show 'SP.
In the lower SV display, the set-point memory SP1 and
additionally LED SET will blink.
Parameter for set-point memory SP1
28
Select one of the four set-point memories SP1, SP2, SP3 or
SP4 using button 5 or 6.
By pressing button P the corresponding set-point memory
will then be opened.
In the upper PV display, the selected set-point memory e.g.
SP3 will blink.
On the lower SV display, the corresponding currently-select-
ed set temperature is displayed.
Input in set-point memory SP3
By pressing button 5, the set temperature is increased.
By pressing button 6, the set temperature is decreased.
By pressing button 5 or 6 the value is
increased or decreased by 0.01 °C. If the
keys are held down for at least 1 sec.,
however, the value increases or decreases
more rapidly, and after 2 secs. even faster,
so that the desired value can be reached
very quickly.
By pressing button P, the newly-set temperature is accepted.
The set-point memory is left and the screen returns to the
parameter level.
To return to calibration mode, press button 6 or button 5
for a long time.
If no button has been pressed for approx.
15 secs., an automatic return to one level
above in calibration mode occurs.
WIKA operating instructions, models CTD9100, CTB9100, CTM9100