Herunterladen Inhalt Inhalt Diese Seite drucken

WIKA CTD9100 Betriebsanleitung Seite 31

Temperatur-blockkalibrator, mikrokalibrierbad, multifunktionskalibrator
Vorschau ausblenden Andere Handbücher für CTD9100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7. Operating elements of the calibrator/micro calibration bath
If button P is pressed again, this opens the parameter
level.
The upper PV display will show 'rEG.
In the lower SV display, SLor will blink.
Parameter for heating gradient, SLor
7.4.8.1 Setting the heating gradient
The heating gradient, SLor, is active if the reference temper-
ature is lower than the set temperature.
The setting range stretches from 99.99 °C/min to 0.00 °C/min.
CAUTION!
The function is deactivated if SLor = InF (in
no Function) is set.
You are at the parameter level (as described in chapter
7.4.1 "Menu structure, parameter levels").
The upper PV display will show 'rEG.
In the lower SV display, SLor will blink.
Parameter for heating gradient, SLor
Press button P.
On the upper PV display, SLor will blink.
On the lower SV display, the corresponding currently-
selected heating gradient is displayed.
Input of the heating gradient
WIKA operating instructions, models CTD9100, CTB9100, CTM9100
By pressing button 5, the heating gradient, SLor, will be
increased.
By pressing button 6, the heating gradient, SLor, will be
decreased.
By pressing button 5 or 6 the value is
increased or decreased by 0.1. If the keys
are held down for at least 1 sec., however,
the value increases or decreases more
rapidly, and after 2 secs. even faster, so
that the desired value can be reached very
quickly.
By pressing button P, the newly-set heating gradient, SLor,
is accepted.
The display returns to the parameter level and the other
parameters can be set.
CAUTION!
If no button has been pressed for approx.
15 secs., an automatic return to one level
above in calibration mode occurs.
After the setting has been made, the calibra-
tor or the micro calibration bath will not use
the new values until you either change the
set temperature or the calibrator or the micro
calibration bath has been turned off and
then on again.
7.4.8.2 Setting the cooling gradient
The cooling gradient, SLoF, is active if the reference temper-
ature is higher than the set temperature.
The setting range stretches from 99.99 °C/min to 0.00 °C/min.
CAUTION!
The function is deactivated if SLoF = InF (in
no Function) is set.
You are at the parameter level (as described in chapter
7.4.1 "Menu structure, parameter levels").
The upper PV display will show 'rEG.
In the lower SV display, SLor will blink.
Parameter for heating gradient, SLor
GB
31

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Ctb9100Ctm9100

Inhaltsverzeichnis