6.
Use the
and confirm with
(non-flashing value). You can cancel the procedure with
7.
Repeat these steps if you wish to get all four calibration points. Use a different standard
foil for each calibration point.
8.
Press
to exit the multi-point calibration.
5.1.3
Delete calibration
To start a new calibration, you can delete the calibration points
recorded during zero-point calibration and during multi-point
calibration individually. Follow these steps:
to enter the "Root" menu.
1.
Press
2.
Use the
with
.
3.
Select according to the following subsections (a, b, c) what
values you would like to delete.
a)
Delete multi-point calibration values
Navigate to "Point Cal" by using the
1.
2.
Use the
confirm with
3.
A dialogue box opens. Confirm with
procedure with
b)
Delete zero-point calibration values
1.
Use the
2.
Use the
confirm with
3.
A dialogue box opens. Confirm with
procedure with
c)
Delete all calibration values
1.
Use the
2.
A dialogue box opens. Confirm with
procedure with
5.2
Measurement
Procedure:
1.
First connect the desired probe as described in chapter 4.2.
2.
Make the required device settings (measure mode, group mode, probe mode, speed,
etc.) as described in chapter 4.3.
3.
Make a zero-point calibration and a multi-point calibration before the measurement
(chapter 5.1).
4.
Place the probe vertically on the surface to be measured. Lift the probe when you
hear the beep sound.
keys to adapt the reading to the actual foil thickness (flashing value)
. The calibration point will then be displayed in the dialogue box
keys to navigate to "Calibration" and confirm
keys to navigate to "Clear Magnetic" or "Clear Magnetic" as required and
.
.
keys to navigate to "Zero Cal" and confirm with
keys to navigate to "Clear Magnetic" or "Clear Magnetic" as required and
.
.
keys to navigate to "Clear All" and confirm with
.
© PCE Instruments
.
keys and confirm with
to delete the calibration values or cancel the
to delete the calibration values or cancel the
to delete all calibration values or cancel the
21
.
.
.