Device status
The device status is indicated by the status LED (Figure 1), as
shown in the table below.
State
Indication
Alarm
Steady red LED
Powered [1]
Single flashing red LED
Fault or Dirty Optical
Double flashing red LED
Chamber [2]
[1] This indication can be disabled using the NC-D-PT Handheld
Programmer.
[2] See "Maintenance and testing" below for details on how to clean or
replace the optical chamber.
Configuration
Configure the detector sensitivity and status LED powered
indication using the NC-D-PT Handheld Programmer.
To connect to the detector:
1.
Connect the 3.5 mm jack of the cable to the BUS
connector on the handheld programmer.
2.
Connect the two crocodile clips to terminals 1 and 3 on the
underside of the detector head (see Figure 2, item 1).
To configure the detector settings:
1.
Connect the programmer to the detector, and then turn the
programmer on.
The LCD displays H003 to indicate that the programmer is
powered up and ready to use.
2.
Enter the Unlock password 456, and then press Clear.
3.
Press Function, and then press 3.
The LCD displays –.
4.
Enter the required setting (see Table 1), and then press
Program.
The LCD displays P to indicate that the setting is
configured.
5.
Press Clear to clear the screen.
6.
Enter the Lock password 789, and then press Clear to lock
the programmer.
The available detector settings are shown below. The default
setting is 1.
Table 1: Optical/heat detector settings
Setting [1]
Description
Normal smoke sensitivity
1 (default)
Heat Class A2R (62°C)
2
Low smoke sensitivity
Heat Class A2R (62°C)
129
Normal smoke sensitivity
Heat Class A2R (62°C)
130
Low smoke sensitivity
Heat Class A2R (62°C)
[1] Only settings 1 and 129 comply with EN 54-7. All settings comply
with EN 54-5.
2 / 25
www.acornfiresecurity.com
See the NC-D-PT Handheld Programmer User Guide for more
information on how to configure or read detector settings.
Maintenance and testing
Maintenance and cleaning
Basic maintenance consists of a yearly inspection. Do not
modify internal wiring or circuitry.
Clean the outside of the detector using a damp cloth. If
required, clean the inside of the detector (to remove dust and
debris) using a vacuum cleaner.
Cleaning the optical chamber
Clean the optical chamber periodically to minimise nuisance
alarms.
To clean the optical chamber:
1.
Remove the detector cover by gently prising the four clips
on the underside of the detector with a screwdriver (see
Figure 2, item 1).
2.
Remove the optical chamber by gently prising each side of
the chamber with a screwdriver (see Figure 2, item 2).
3.
Clean the optical chamber using compressed air.
4.
Replace the optical chamber, and then replace the
detector cover (ensuring that the status LED is aligned
with the hole in the detector cover).
Always test the detector after cleaning the optical chamber.
Testing
Always test detectors after installation to ensure that the
detector communicates it's status to the control panel.
The detector must be powered up for at least 10 seconds
before testing.
To perform a magnetic test:
1.
Hold a magnet close to the test area of the detector head
(see Figure 1) until the status LED indicates an alarm.
After testing, reset the control panel to reset the status LED.
Specifications
Electrical
Operating voltage
Current consumption
Standby
Status LED
Alarm
Power indication on
Ripple voltage
Polarity sensitive
Power indication on
Remote indicator alarm output
Power indication off
Power indication off
www.acornfiresecurity.com
9.5 to 28 VDC
60 µA at 24 VDC
55 mA at 24 VDC
2 V max. (peak to peak)
Yes
2 mA max.
P/N 3103029-ML • REV 001 • ISS 04JUN24