Harmonic distortion
Voltage fluctuations/flicker
) The device is intended for use in a home environment or professional healthcare structure and may exclusively
a
be used in hospital rooms and in sections of hospitals or clinics set up for respiratory therapy. The strictest
acceptance limits required for Group 1 Class B (CISPR 11) have been considered and applied. The device is
suited for use in the environments mentioned above so long as it is connected to the public electric mains.
) The test may be applied to this environment so long as the ME EQUIPMENT and ME SYSTEM used are
b
connected to the PUBLIC ELECTRIC MAINS and the power supply is compliant with the standard applicable
policy regarding electromagnetic compatibility (EMC).
Manufacturer's Declarations and Guidelines - Electromagnetic immunity - Enclosure Door
Phenomenon
ELECTROSTATIC
DISCHARGE
Radiated radio-frequency
electromagnetic fields
Fields near radio-
frequency devices for
wireless communication
Electromagnetic fields for
RATED mains frequency.
a
) The device is intended for use in a home environment or professional healthcare structure and may
exclusively be used in hospital rooms and in sections of hospitals or clinics set up for respiratory therapy. The
strictest limits for acceptance were considered and applied during immunity testing.
) Before modulation.
b
c
) This test level presupposes a minimum distance of at least 15 cm between the AirPremiumSystem
02038201100000 and the sources of magnetic fields at the mains frequency.
Manufacturer's Declarations and Guidelines - Electromagnetic immunity - Door for access with
Phenomenon
Fast electrical transient/
burst
Applicable EMC policy
or test method
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
alternating current
Applicable EMC policy
or test method
IEC 61000-4-4
18
IEC 61000-3-2
COMPLIANT WITH IEC 61000-3-3
Immunity test levels
Professional
healthcare
environments
± 8 kV in contact
± 2 kV, ±4 kV ±, ±8 kV, ±15 kV air
)
a
COMPLIANT
NB: further information about the distances to
be maintained from portable and mobile radio-
frequency communication devices (transmitters)
and AirPremiumSystem
be requested by contacting PIKDARE S.p.A. using
the contact information included in this booklet.
It is nevertheless advisable to keep the electrome-
chanical aerosol therapy device at a safe distance
(at least 1 m) from mobile phones or other com-
munication devices that use radio frequencies, to
minimise interference.
30 A/m
)
c
50 Hz or 60 Hz
Immunity test levels
Professional
healthcare
environments
± 2 kV
Repetition frequency at 100 kHz
) Class A
b
)
b
Home environment
10 V/m
)
b
80 MHz - 2.7 GHz
80% AM at 1 kHz
02038201100000 may
Home environment
Inhaltsverzeichnis
loading

Verwandte Produkte für Pic Solution AirPremium System

Diese Anleitung auch für:

02038201100000

Inhaltsverzeichnis