Scarce nebulisation
Nebulisation absent
Nebulisation absent
Slow nebulisation
TECHNICAL SPECIFICATIONS
Nominal voltage: 230 V~
Power: 170 VA
Maximum flow: 18±2.5 l/min
Max pressure: 3,2 ± 0,4 atm
Conditions for using the device: CONTINUOUS
MMAD: 2.28±0.04 µm
Nebuliser cup capacity: 10 ml
Appliance service life: 3,200 treatments
Service life of accessories: 5 years
Service life in storage conditions: 1 year
*Data on noise levels were measured on a new device. Values may vary with use.
The information on the unit's capabilities provided by the manufacturer in accordance with EN 13544-1 may not
apply to medicines provided in suspension form or those that are highly viscous.
EN 60601-1 Medical electrical equipment – Part 1: general requirements for basic safety and essential performance.
EN 60601-1-2 Medical electrical equipment – Part 1: general requirements for basic safety and essential performance.
- Collateral standard: electromagnetic compatibility - Requirements and tests.
EN 13544-1 Respiratory therapy equipment - Part 1: Nebulising systems and their components.
This section contains information specific to product compliance with the EN 60601-1-2 standard. AirPremiumSystem
is a medical electrical device that requires special precautions regarding electromagnetic compatibility and needs to be
installed and commissioned according to the electromagnetic information provided.
Mobile and portable RF communications equipment (mobile phones, transceivers, etc.) may affect the appliance.
AirPremiumSystem
02038201100000 is designed for use in an electromagnetic environment that meets the
specifications provided below. The person who purchased or who uses AirPremiumSystem
responsible for ensuring that the device is used in an environment that complies with said specifications.
Manufacturer's Declaration and Guidelines – Electromagnetic Emissions
Phenomenon
Radiated and conducted radiofrequency EMISSIONS
Nebuliser cup clogged
Nebuliser cup components not
assembled correctly
The compressor does not work
Drug too thick
Frequency: 50 Hz
Fuse: T 1,6A L 250 V~
Operational flow: 8 ± 1.5 l/min
Operational pressure: 1,4± 0,2 atm
Weight: 2.2 kg
Breathable fraction: 89,2±0,43%
Noise level at 50 cm*: 72 dB
Professional healthcare
environments
)
a
17
If the problem persists after clean-
ing, replace the nebuliser cup.
Take apart and reassemble the nebu-
liser cup; If the problem persists, re-
place the nebuliser cup.
Be sure that the plug for the power
cord is disconnected from the
socket, check that the on/off switch
is in the "on" position and that the
electrical socket works.
Dilute the medicine as indicated
by the pharmaceutical company
02038201100000 is
HOME
)
ENVIRONMENT
a
CISPR 11
Group 1 Class B
)
a