NOTE 1: At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electro-
magnetic propagation is affected by absorption and reflection from
structures, objects and people.
Guidance and manufacturer´s declaration - electromagnetic
immunity. The device is intended for use in the electromagne-
tic environments listed below, and should only be used in such
environments:
Immunity
test
Radiated
RF
IEC 61000-
4-3
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Elec-
tromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
y BA.indd 25
IEC
Com-
Electromagnetic environment -
60601-
pliance
guidance
test
level
level
Portable and mobile RF communications
3V/m
3 V/m
equipment should be used no closer to any
80 MHz
part of the device, including cables, than
to 2.5
the recommended separation distance
GHz
calculated from the equation applicable to
the frequency of the transmitter
Empfohlener Trennungsabstand:
d= 1.2 √P
d = 1.2√P
80 MHz to 800 MHz
d = 2.3√P
800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,a should be less than the compli-
ance level in each frequency range.b