POSSIBLE PROBLEMS AND SOLUTIONS

Problem
SpO
cannot be shown
normally.
SpO
shown
unstably
The oximeter
Cannot be
switched
On.
„Finger out"
Displayed on
screen
Indication lamps
suddenly
go off
"Error3" or
"Error4"
Displayed on
screen
"Error7"
displayed
on screen
y BA.indd 21
Possible Reason
or PR
1.
Finger is not
2
inserted correctly.
2.
Patient's oxyhaemoglo-
bin level is too low
to be measured
or PR is
1.
Finger might not be
2
inserted far enough.
2.
Finger is trembling or
the patient's body is
moving.
1.
Battery power may be
inadequate or not be
there at all.
2.
Batteries may have
been inserted incor-
rectly.
3.
The oximeter might be
damaged.
1.
There is no finger on
the pulse oximeter or
finger is not positioned
properly.
1.
The product is automa-
tically switched off if no
signal is detected for
longer than 8 seconds.
2.
Low power/ Batteries
empty.
1.
Low power/ Batteries
empty
2.
Receiving sensor
shielded or
sensor damaged.
1.
Low power/ Batteries
empty
2.
Infrared sensor
damaged.
Solution
1.
Try inserting the finger
again.
2.
Try several more times
if you are sure that there
is no problem with the
device. Please go to a
hospital in good
time to obtain a precise
diagnosis.
1.
Try inserting the finger
again.
2.
Try not to move.
1.
Please replace batteries.
2.
Please reinsert the
batteries.
3.
Have the device tested by
the manufacturer
1.
Place your finger in the
oximeter.
1.
Normal
2.
Replace the batteries.
1.
Insert new batteries.
2.
Have the device tested by
the manufacturer.
1.
Change the batteries.
2.
Have the device tested by
the manufacturer.