Short protocol Homogeniser Precellys
Problem
Possible cause
Impaired running
The four feet do not rest on a level
smoothness, in-
surface.
creased vibration
The fan does not
Electric power supply is discon-
work
nected.
Faulty fuse
Faulty cooling system
No display on the
Electric power supply is discon-
screen
nected
Faulty fuse
Faulty display
One or several
The cap is not screwed properly.
tubes are not tight
Sample material or beads sitting
between cap and tube prevent tight
closing.
Tubes are disrupted by the kinetic
energy of the beads.
Tubes are disrupted by gas pres-
sure.
Power failure, pull-
For safety purposes the vacuum
ing the main plug
between indented plate and sam-
during the run.
ple holder is held.
PEQLAB_v0314_E
®
24 / 24-Dual
25
Suggestions
Adjust the feet with the help of a screw wrench. Please
work with someone together to carefully tilt the Precel-
®
®
lys
24/Precellys
24-Dual.
®
Never set the Precellys
24/Precellys
because the weight could damage the casing.
Check the unit is plugged in properly. Make sure that
the power input filter on the back of the device
matches the voltage range delivered by your local
mains supply.
Replace fuse.
Turn off the system and contact TECHNICAL SUPPORT
at PEQLAB.
Check the unit is plugged in properly. Make sure that
the power input filter on the back of the device
matches the voltage range delivered by your local
mains supply.
Replace fuse.
Turn off the system and contact TECHNICAL SUPPORT
at PEQLAB.
Lock the tubes properly. Clean and/or disinfect the
system as described in 'CLEANING AND DECON-
TAMINATION'.
Occurs eventually using very small beads (e.g. VK01).
Pay attention that there is no sample material or beads
located at appropriate sites of the cap or tube (option-
ally tubes can be briefly centrifuged).
Use exclusively high-quality screw cap tubes e.g. the
Precellys lysis kits of PEQLAB. By using the larger
beads (e. g. CK28 or MK28) the sample volume
should be at minimum 500 μl (see also 'General user
tips')!
The maximum sample volume of 1.4 ml in the 2 ml
tubes and 5 ml in the 7 ml tubes should not be ex-
ceeded in order to guarantee sufficient compressible
volume for a potential heat expansion in the tube.
Make sure the electrical power supply is working and
turn the system back on. The vacuum is relieved then
and the indented plate can be removed.
®