Short protocol Homogeniser Precellys

Combinable parameters/Refractory period

For safe operation the system is allowing the following combinations of program parameters.
Refractory periods are rest times in which a consecutive run cannot be started. They prevent the system
from overheating. However, the risk of overheating only exists during extensive high throughput appli-
®
cations using Precellys
which will shut off the motor if the temperature rises above the limit. Therefore, from firmware version
4.1 (12/2011) on refractory periods do not exist any more in Precellys
by pressing 'cancel' in Precellys
The actual firmware version of the system is displayed when pressing 'cancel' for 10 sec after switching
on the unit.
Repetition of an agitation
Speed [rpm]
2500
4000 - 6100
6200
6300 - 6500
*
6600 – 6800

Instructions for the use of 7 ml tubes in the Precellys

To prevent tube breakage caused by thermal expansion please comply to the following program pa-
rameters if using the 7 ml tubes in the Precellys
Repetition of an agitation
Speed [rpm]
2500
4000 - 5500
5600 - 6000
6100 - 6500
PEQLAB_v0314_E
®

24 / 24-Dual

24. In addition system temperature is always monitored by a temperature sensor
®
24.
interval
1 – 3
1 – 3
1 – 3
1 – 3
1 – 3
Duration of an interval
interval
1 – 3
1 – 3
1 – 3
1 – 2
Duration of an inter-
val [s]
5 – 90
5 – 90
5 – 90
5 – 30
31 – 60
5 – 30
®
24-Dual
®
24-Dual.
[s]
5 – 60
5 – 40
5 – 30
5 – 20
19
®
24-Dual and can be skipped
Pause between
Refractory period*
intervals [s]
5 – 120
5 – 120
5 – 120
5 – 120
15 – 120
30 – 120
* Precellys
2 min
2 min
5 min
5 min
5 min
®
24 only