Herunterladen Diese Seite drucken

Consort R735 Anleitung Seite 6

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1. Press SET until the display shows [SEt rEL] and press CAL.
2. The display shows e.g. [r1=PH.Hi], [r1=PH.Lo], [r1=ALrM]...etc. Select the
desired mode for relay-1 and press CAL each time.
3. The display shows the delay (s) after which the relays should be activated,
when one of the pre-set levels is exceeded, e.g. [r1.dt=120]. Pre-set the
desired delay for the relays and press CAL.
1. Press SET until the display shows [SEt ALArM] and press CAL.
2. The display shows e.g. [ALrM PH], [ALrM MV], [ALrM EC]...etc. Select the
desired mode and press CAL.
3. The display shows [AL.= on] or [AL.= OFF]. Select the desired mode and press
CAL.
4. The display shows e.g. [AL.=6.00][LoW]. Select the desired minimum level and
press CAL.
5. The display shows e.g. [AL.=8.00][HiGH]. Select the desired maximum level
and press CAL.
6. The display shows the hysteresis between the ON and OFF state of the relays
e.g. [AL.=0.10][HySt]. Pre-set the desired hysteresis and press CAL.
1. Press SET until the display shows [SEtUP] and press CAL.
2. The display lets you choose between [bEEP on] or [bEEP OFF]. Decide whether
or not the beeper should sound each time the keyboard is touched. Press
CAL.
3. The display shows [ScAn OFF] or [ScAn on]. Decide whether the display should
alternate between the selected modes with an interval of 4 s or not and press
CAL. Pressing PRINT in this mode will print all values of the selected chan-
nels.
4. The display shows e.g. [br=2400]. Pre-set the desired baud-rate (150...4800
b/s) and press CAL.
5. The display shows e.g. [rS=120]. Select the desired time interval (0...9999 s)
between the transmitted data (RS232). Pre-set to zero if no automatic printing
is required. Press CAL.
6. The display shows e.g. [Id.no=5]. Select the desired identification number
(0...999) for your meter and press CAL. The instrument can be identified by
e.g. a computer when a specific number is allocated to it. Pre-set to zero to
ignore this number.
7. The display shows [Code on], [Code OFF] or [Code CAL]. Decide whether all
future access to the instrument should be by entering a code (on) or not (off).
Press CAL. Select [Code CAL] to enter your personal code to avoid undesired
access to the instrument. The display shows [E=-----] while [=] is blinking.
Enter your secret sequence of 5 keys to proceed. Press CAL.
1. Select the mV range by pressing MODE. The display will immediately show the
measured mV according to the previous calibration. Should you want to recali-
brate, press CAL.
2. Immerse the electrodes in a standard solution of known potential.
3. The display shows e.g. [MV=476.3]. Adjust to the proper value and press CAL
again or press RES to reset the absolute readings.
4. After rinsing the electrodes with distilled water, immerse them in the solution
to be measured and read the display.
5. Rinse the electrodes always with distilled water after use and store them in a
3...4 M KCl solution.
R735 èwww.consort.be
Relais
Alarm
Specific features
mV measurement
6

Werbung

loading

Verwandte Produkte für Consort R735