Maintenance
7 Maintenance
7.1 General maintenance information
Danger to life from electric shock during maintenance and service work
The device is only completely de-energized when the mains plug has been disconnected and the tur-
bopump is at a standstill. There is a danger to life from electric shock when making contact with live
components.
► Before performing all work, switch off the main switch.
► Wait until the turbopump comes to a standstill (rotation speed f = 0).
► Remove the mains plug from the device.
► Secure the device against unintentional restarting.
Health hazard through poisoning from toxic contaminated components or devices
Toxic process media result in contamination of devices or parts of them. During maintenance work,
there is a risk to health from contact with these poisonous substances. Illegal disposal of toxic sub-
stances causes environmental damage.
► Take suitable safety precautions and prevent health hazards or environmental pollution by toxic
► Decontaminate affected parts before carrying out maintenance work.
► Wear protective equipment.
Risk of cuts on moving, sharp-edged parts when reaching into the open high vacuum connec-
tion
Incorrect handling of the turbopump before maintenance work results in hazardous situations with
risk of injury. There is a risk of cuts from accessing sharp-edged, rotating parts when removing the
turbopump.
► Wait until the turbopump comes to a standstill (rotation speed f = 0).
► Switch the turbopump off properly.
► Secure the turbopump against re-start.
► Seal open connections immediately following removal, using the original protective cover.
7.2 Maintenance intervals and responsibilities
Recommendations for performing maintenance measures
1. Clean the turbopump exterior with a lint-free cloth and a little isopropanol.
2. Replace the electronic drive unit as an independent unit.
3. Replace the operating fluid reservoir as an independent unit.
4. Pay attention to when the operating fluid must be changed.
5. Change the operating fluid reservoir at least every 4 years.
6. Have Pfeiffer Vacuum Service replace the rotor bearing of the turbopump at least every 4 years.
7. Consult with Pfeiffer Vacuum Service about shorter maintenance intervals for extreme loads or im-
8. For all other cleaning, maintenance or repair work, contact the appropriate Pfeiffer Vacuum Serv-
102/126
WARNING
WARNING
process media.
WARNING
pure processes.
ice location.