Basic calibration after exchange of electronic module
The antenna and the electronic module (electronics + HF module) are system-specific. The exchange of a different type of electronic
module would cause a change in the measuring features of the instrument. Therefore it is important to perform a basic calibration on the
instrument after an exchange of electronic module.
The exchange of electronic module is only allowed for spare part with identical features!
Preparation for level calibration:
Customer specific parameters can be stored by means of up / down loads. A basic calibration must be performed in any case.
Perform basic setup according to operating manual.
For operation calibrated to legal standards the level
is controlled and corrected manualy with a dip table
see parameter 009 "setting".
If necessary, record a new dip table
in group selection "dip table", see operating manual.
For operation calibrated to legal standards (according to OIML R85),
access to the electronics must be locked with the lock switch,
which has to be sealed by an official.
SD
E
F
2
2
Instrument is assembled and
mounted on tank
End
when using a Tank Side Monitor,
further compensation is possible
(i.e. pressure, temperature, etc.)
E = zero (empty)
adjusted in 005
D
F = span (full)
adjusted in 006
D = distance (between flange / level)
displayed in 0A5
L = level
displayed in 0A6
L
SD = safety distance
adjusted in 015
Exchange of electronic module on Micropilot S FMR540