1. CONVENTIONAL USE
The equipment may only be used for the purpose as set forth in the catalogue and in the technical descrip-
tions which means for ultrasonic cleaning as described in this instruction, and only in conjunction with
external devices and components which are recommended or approved by EMAG AG. Any other use of the
equipment is considered to be non conventional.
In case of non conventional use, EMAG AG will not be liable for possible damages and will not warrant for
the correct and functional work of the components. Part of the conventional use is also the observance of
all information given in the operating manual. To this attention is drawn by the
The correct and safe operation of the equipment implies appropriate transportation, storage, installation
and assembly as well as careful handling and maintenance.

2. NON CONVENTIONAL USE

In case of non conventional use of the equipment danger for life and limb as well as danger for property
may occur. Therefore a non conventional use is prohibited. In case of non conventional use EMAG AG
will not be liable for possible damages and will not warrant for the correct and functionally work of the
components. In order to avoid non conventional use of the ultrasound cleaning equipment the following
directions are to be considered:
• The equipment may only be operated with the voltage printed on the type plate on the backside/ bot-
tom side of the housing (shockproof socket).
• No flammable or aggressive chemicals (i.e. acids etc.) must be used for operation of the equipment. If
necessary, use glass container EM 50 DG.
• During operation do not touch the liquid.
• In case of transportation damage, don`t connect the equipment with main source.
• Do not clean creatures (i.e. finger nails, animals etc.) in the equipment.
• The equipment must only be operated with the correct level of water (water till mark, approx.
1 cm below upper brim). This must be controlled if equipment is operated for a longer time.
• Strongly contaminated goods shall be positioned in the equipment with the contaminated side to the
bottom – do not staple goods.
• The goods to be cleaned must not be positioned directly onto the bottom of the tub – use the basket.
• If the equipment is operated for a longer time, do not stay longer than 4 hours within 1 m distance to the
equipment.
• The equipment must only be operated indoor and within a dry environment.
• It is normal, that the upper brims warm up at longer operation.
• Before emptying the tub, the power connector must be disconnected from the main source.
3. OPERATING INSTRUCTION
Check the specification table fort he correct operating requirements (located on the rear of the unit).
Startup
Before connecting power, put the unit on a stable and flat working surface. Add water to tank, suitable to
completely immerse the items to be cleaned.
Connect the power cable on the rear of the unit and ensure the wall supply is switched „ON". Touch „ON/
OFF" button on the front panel to turn on the machine. You are now ready to set the timer and temperature.
8
sign on the type plate.