Guidance and manufacturer's declaration-electromagnetic immunity
The Thermometer is intended for use in the electromagnetic environment specified below. The
customer or the user of the Thermometer should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Compliance
IEC 60601 test level
level
±6kV
±6kV contact
contact
±8kV air
±8kV air
3 A/m
3A/m
12 | GB
Electromagnetic environment-
guidance
Floor should be wood, concrete or
ceramic tile. If floors are covered
with shythetic material, the relative
humidity should be at least 30%.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.