• Only process media that will not react dangerously to the extra en-
ergy produced through processing. This also applies to any extra
energy produced in other ways, e.g. through light irradiation.
Beware of the risk of:
DANGER
- flammable materials
- glass breakage
- sharp edges on the dispersing tool.
• Process pathogenic materials only in closed vessels under a suit-
able fume hood. Please contact IKA if you have any questions.
• Abrasion of the dispersion equipment or the rotating accessories
may get into the medium you are working on.
• Powder can not be too close to the flange. It can be blown away
by air turbulences of the drive.
• The instrument starts up again following a cut in the power supply.
Please note that the instrument must be switched off at the ON/
OFF switch in this case before the power supply is reconnected.
• The instrument can only be disconnected from the mains supply
by pulling out the mains plug or the connector plug.
• The socket for the mains cord must be easily accessible.
• Always disconnect the plug before fitting accessories.
• Safe operation is only guaranteed with the accessories described
in the "Accessories" chapter.
Never run dispersion tools dry, as the gasket
WARNING
and bearings will be destroyed if the tools are
not cooled by the medium.
• The voltage stated on the nameplate must correspond to the
mains voltage.
• Protect the instrument and accessories from bumps and impacts.
• The instrument can only be opened by experts.
• Removable parts must be refitted to the instrument to prevent
the infiltration of foreign objects, liquids etc.
• Do not cover the ventilation slots on the drive in order to ensure
adequate cooling of the drive.
• The gasket and bearings are made of PTFE and rustproof steel; the
following points should therefore be noted: Chemical reactions of
PTFE occur in contact with molten or solute alkali metals and alka-
line earth metals, as well as with fine powders of metals in groups 2
and 3 of the periodic system at temperatures above 300 °C - 400 °C.
Only elementary fluorine, chlorotrifluoride and alkali metals attack
it; halogenated hydrocarbons have a reversible swelling effect.
(Source: Römpps Chemie-Lexikon and "Ulmann", Volume 19)
Intended use
Use:
When used in combination with one of our recommended dispersing
elements, the drive unit is a high-speed dispersing and emulsifying unit
capable of handling free-flowing and liquid media in batches.
- Production of:
Emulsions
Dispersions
Wet crushing
- Operating modes:
Hand hold mode
On stand
Range of use:
Indoor environments similar to that a laboratory of industry area.
The safety of the user cannot be guaranteed:
- If the instrument is operated with accessories that are not supplied
or recommended by the manufacturer
- If the instrument is operated improperly or contrary to the manu-
facture's specifications
- If the instrument or the printed circuit board are modified by the
third parties.