8 Protective measures and warnings
Please observe also the instructions provided on working with the neutral electrode in
chapter 6.5, the operating instructions with the accessories and the applicable safety
measures!
!
Before any procedure, check the cables, accessories (handles,
electrodes, forceps) and foot switches for any signs of visible
irregularity. Do not use cables or instruments with broken or
defective insulation. These can result in burns if they come into
The unit should only be operated at a mains socket with protective earth contact.
When performing electrosurgical procedures, the risk of burns must be kept as small as possible.
For this, the following safeety measures must be observed:
!
All plug-in connections may only be carried out with compatible accessories.
!
Use the greatest care when working with the neutral conductor!
!
Protect the patient from contact with earthed metallic elements (OP table, infusion stand
etc.) during use.
!
Avoid contact between supply cables and patients, other cables or instruments! Ensure that
there are no loops in the cables of handles and electrodes or fixture tot he OP table.
!
Avoid small-area skin-to-skin contact in the patient (e.g. a finger making contact with the
thigh).
!
Avoid contact between HF instruments and non-insulated instruments.
!
Prevent the penetration of fluids between the patient's body and the neutral electrode.
!
The output power should not be set higher than is absolutely necessary fort he procedure.
!
Do not touch the patient during activation oft he coagulation or cutting current. Should this
be necessary, a piece of fabric can be used for insulation purposes.
!
In patients connected to monitoring devices (ECG), the monitoring electrodes must be
attached at the greatest possible distance to the application areas oft he electrosurgical
electrodes.
contact with the skin!