6 Starting Up
6.1.1 Climatic conditions
The following requirements must be fulfilled with respect to the climatic conditions at the place of use of the
pyrometer:
• temperature range 0 °C to +70 °C
• maximum air humidity 60 % at room temperature
• Atmosphere free of dust, corrosive vapours, and fume gases
6.1.2 Requirements at the place of use
Take into account the following requirements at the place of use of the pyrometer:
• Take care with the choice of the place of use and take into account the ergonomic and the legal guidelines
for industrial safety in order to ensure safe operation of the pyrometer.
• Set up the pyrometer on a firm, stable base. The base must be free from concussion and vibration.
Note:
We recommend using the rail mounting plate available as an accessory for the attachment of the pyrom-
eter as well as a clamping attachment (see. Accessories (section 3.3) at page 4.
• Do not expose the system to direct sunlight or radiation from heating (e.g. radiators).
• Do not set the pyrometer near strong electromagnetic fields or near electromagnetic interference sources.
• Make sure that the pilot light falls unhindered on the object being measured and that there is nothing in
the path of the rays.
6.2 Installation
DANGER:
There is danger of injury and possible equipment damage by connection of cables under power!
Never connect cables under power! Make sure that the voltage supply is switched off before connection
of the cables to the pyrometer!
IMPORTANT:
Operation of the pyrometer about ambient temperatures of more than 70°C is only permitted with water
cooling! Connect the pyrometer to a suitable cooling circuit with ordinary tap water! The use of coolants
is permitted only with the agreement of the manufacturer!
The temperature of the cooling water must lie between 10 °C and 35 °C. Condensation build-up due to
the low temperature of the cooling water is to be avoided.
For operation the switch a power supply of 24 V DC
power supply ensure correct polarity. To switch off the instrument, unplug the connector or interrupt the power
supply.
To meet the electromagnetic requirements (EMI), a shielded connecting cable must be used. The shield of
the connecting cable has to be connected only on the pyrometer's side. On side of the power supply (switch
board) the shield must be open to avoid ground loops.
The switch does not require time for starting or preheating and is immediately ready for operation.
The connecting cable (please order it separately) has a 3-pin plug at one end, wires on the other.