15 Modbus Kommunikations-Protokoll
Table 1 Read-only register address list (read with 0x03 function code)
Channel
description
Velocity
Flow
Totalizer (float)
Totalizer
(integer part)
Totalizer
(decimal part)
Daily totalizer
(integer part)
Daily totalizer
Monthly totalizer FLOAT
Yearly totalizer
4-20mA output
Running time
Meter serial
number
Date and time
Signal Quality Q UINT
Running status
Meter Address
Communication
baud rate*
Velocity unit
Flow unit
Totalizer unit
S462
Format
Length
FLOAT
4-byte
FLOAT
4-byte
FLOAT
4-byte
UINT32
4-byte
FLOAT
4-byte
UINT32
4-byte
FLOAT
4-byte
4-byte
FLOAT
4-byte
FLOAT
4-byte
UINT32
4-byte
String
8-byte
String
6-byte
2-byte
UINT
2-byte
UINT
2-byte
UINT
2-byte
STRING
4-byte
STRING
4-byte
STRING
4-byte
Register
Description
address
0x0000
Unit: m/s
0x0002
Unit: as programmed in m3/h,
l/min or GAL
0x0004
0x0006
0x0008
Unit: as programmed in m3, liter
0x000A
or GAL
0x000C
0x000E
0x0010
0x0012
Unit: mA
0x0014
Unit: sec
0x0016
N/A
0x001A
Year, month, day, hour, minute,
second
0x001D
0 ... 99
0x001E
0 = normal
1 = abnormal
0x001F
Slave address 1 ... 249
0x0020
0 =2400,1 = 4800, 2 = 9600,
3 = 19200
0x0021
"m/s" or "f/s"
0x0023
0x30 = m3/h
0x31 = LPM
0x32 = GPM
0x0025
0x30 = m3
0x31 = L
0x32 = GAL
25