How to carry out adjustments
 Switch off weighing system
 Press
for 30 sec., PAR is displayed
 Press
until CAL is displayed
 Press
, ZERO is displayed flashing, because the
reference height is not reached yet
 Place a pallet and adjust the reference height, when the
reference height is reached, the display stops flashing and
only the right-hand bar above in the display continues to be
displayed.
 Press
for approx. 3 sec., the count down fur setting
the first adjustment point runs automatically.
The weight of 1000 kg for the first adjustment point is
displayed. One point (=1. adjustment point) is displayed
before the cipher.
 Place 1000 kg
After placing the weight the weighing system leaves the
reference height, the display flashes
 Bring the weighing system on reference height, the display
stops flashing