Herunterladen Inhalt Inhalt Diese Seite drucken

Symbols; Environment; Specifications - CareFusion Smoke Check Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Smoke Check:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Symbols

Type B device
In accordance with Directive 93/42/EEC
0086
Disposal in compliance with WEEE

Environment

This instrument complies with directive EN60601-1-2
electromagnetic compatibility but can be affected by cellular phones
and by electromagnetic interference exceeding levels specified in EN
50082-1:1992

Specifications

Sensor type
Range
Detected levels:
0 to 6ppm
7 to 10ppm
11 to 20ppm
>20 ppm
Accuracy
Sensitivity drift
Sensor life
Response time
Hydrogen cross sensitivity
Operating temperature
Operating pressure
Electro-chemical fuel cell
0 - 20 ppm
Display:
0-6 ppm with green indicator
7-10 ppm with amber indicator
11-20 ppm with red indicator
20+ ppm with red indicator + alarm
+/-5% of reading or 1ppm whichever
is the greater
0.5%/°C
2 to 5 years
< 15 sec (to 90% of reading)
<15%
15 – 25 °C
Atmospheric +/- 10%
14