Maintenance and Service
4. Wash oxygen pipe
Nomally, oxygen pipe shall be washed three times a day. Nosal
head of oxygen pipe shall be washed after each use. Immerse it in
white vinegar for 5 minutes and then wash it with water or clean it
with medical alcohol. It is suggested to replace oxygen pipe once
every two months.
Note: keep the inside of oxygen pipe dry, without any water.

5. Replace fuse

First, cut off the power supply. The fuse holder is located at the
bottom of the machine. Use a cross screwdriver to screw out the
fuse holder cover anticlockwise, and then take it out for
replacement.
(Note: fuse model: T1.6AL250V, Ф5 * 20)
6. Common troubleshooting guide
Fault
phenomenon
The oxygen
generator
haspower
supply, but the
machine does
not work and
the display
does not show
Possible causes
1) Fuse of power supply
is broken
2) Insufficient power at
power outlet
3) If the concentrator is still not working, contact
the equipment supplier or manufacturer
12
Elimination method
1) Replace fuse tube
2) Do not use
extension cords to
move the concentrator
to another power outlet
or circuit