EN
Operating instructions
4. Connection type and diagram
4.1 Output signal Pt100
Connection to constant current source, measuring current 0.3 mA to 1.0 mA.
Round connector M12x1, output signal Pt100
4
3
ϑ
1
2
4.2 Output signal 4 mA ... 20 mA, 2-wire (measuring transducer)
Load resistance diagram:
1183
833
583
0
Voltage L
Round connector M12x1, output signal 4 mA ... 20 mA
+
-
-
+
22
8027569/2022-03-01 © SICK AG • Temperature • Germany • Subject to change without notice
4. Connection type and diagram
1
2
3
4
10
24
in V
+
L
/ Q
+
A
4
3
2
1
M
The permissible load resistance
30
36
depends on the voltage of the loop
supply.
TCT