Herunterladen Inhalt Inhalt Diese Seite drucken

Westfalia 82 04 51 Bedienungsanleitung Seite 58

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6. Read the voltage value from the display. The "-" sign in front of
the reading indicates a negative direction of the measured
current.
Notes:
When the test leads are connected to the unit in idle state (i.e.
without being connected to a device to be measured), a
fluctuating value may be displayed on the display due to the
high sensitivity of the device. When a measurement is per-
formed how-ever, the displayed value corresponds exactly to
the specified accuracy of the unit.
If "OL" is shown on the display at the manual range mode, the
selected measuring range has been exceeded by the meas-
urement value. Select a higher range.
In manual measuring range mode select the highest range
when the value to be measured is unknown.
Do not exceed the maximum input voltage of 600 V
If the measured voltage is higher than 610 V
"OL" appears on the display and the built-in buzzer sounds.
DC Voltage Measurements - Accuracy
Range
660 mV
6.6 V
66 V
600 V
Input impedance:
Overload protection: 660 mV range: 250 V
Max. input voltage:
Operation
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
10 MΩ
66–600 V range: 600 V
600 V
Accuracy
± (0.8% of reading + 3 digits)
± (1.0% of reading + 5 digits)
53
.
, the symbol
or 250 V rms AC
or 600 V rms AC

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis