Herunterladen Inhalt Inhalt Diese Seite drucken

Menu; Error Messages - Velp Scientifica ECO 6 Thermoreactor Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ECO 6 Thermoreactor:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 23

4. Menu

From the main screen, press the DOWN button to access the main menu.
Use the UP and DOWN buttons to browse through the available menus, listed below.
Press the ENTER button to access the selected menu.
Then, use the UP and DOWN buttons to modify the value of the parameter (when it is possible).
Press the ENTER button twice to confirm.
To exit the menu, press the ESC button.
Note: please see the dedicated paragraph for the program setting.
Menu
PROGRAMS
LANGUAGE
TEMPERATURE UNIT
SOUNDS
BLACKOUT
COUNTERS
SERVICE
Programs setting
From the PROGRAMS menu, use the UP and DOWN buttons to browse through the 20 available programs.
Press the ENTER button to access the selected program.
Press the ENTER button to select the program for the analysis.
Press the DOWN button to modify it:
Press the ENTER button
Use the UP and DOWN buttons to modify the working temperature; then press the ENTER button to confirm.
Use the UP and DOWN buttons to modify the analysis time; then press the ENTER button to confirm.

5. Error messages

When the display shows an error message, the instrument's functions are stopped automatically.
To remove the error message, restart the instrument.
If the alarm persists on the display, please contact VELP Scientifica technical service department.
Default value
---
English – Italiano
English
Celsius – Fahrenheit
Celsius
Yes – No
Yes
Yes – No
Yes
---
---
AL2
Temperature probe broken
AL4
Overtemperature
Range
Description
It allows setting the working parameters.
Please see the dedicated paragraph.
It allows setting the software language.
It allows setting the temperature unit.
It allows enabling/disabling acoustic signals.
It allows enabling / disabling the blackout function.
If Yes is selected and a blackout occurs, when the
power supply is restored, the working cycle
restarts automatically.
It shows the number of analyses performed and
the heating element operating time.
Only for authorized technicians.
3

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis