Herunterladen Inhalt Inhalt Diese Seite drucken

AEG BMG 5610 Bedienungsanleitung Seite 133

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
PROBLEM
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er A"
No response when
you press a button
or load batteries.
MAINTENANCE
1.
Do not drop this monitor nor subject it to strong impacts.
2.
Avoid high temperatures and direct sunlight. Do not immerse the monitor in water as this will
result in damage to the monitor.
3.
If this monitor was stored at a temperature near the freezing point, allow it to come to room tempera-
ture before use.
4.
Do not attempt to disassemble this monitor.
5.
If you do not use the monitor for a long time, please remove the batteries.
6.
It is recommended the performance should be checked every 2 years or after repair. Please contact
the service center.
POSSIBLE CAUSE
Cuff pressure above 300 mm Hg
More than 3 minutes with cuff pres-
sure above 15 mm Hg
EEPROM accessing error
Device parameter checking error
Pressure sensor parameter error
Incorrect operation or strong electro-
magnetic interference.
SOLUTION
Measure again after five minutes. If the
monitor is still abnormal, please contact
the local distributor or the factory.
Take out batteries for five minutes, and
then reinstall all batteries.
English
133

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis